The GE Marquette Case System includes the
following:
- Remote viewing function.
- Digital compact acquisition module reduces
non-cardiac electrical noise.
- Waterfall Display color maps all exercise
ECG tests into one quick and easy to read
screen.
- Customizable Stress Chart offers a tabular
summary of all test variables for clear,
consistent and efficient report generation.
Specifications
GE Marquette Case System Performance Specifications
GE Marquette Case Instrument type:
Unity-enhanced cardiac stress testing system
with 14 channel acquisition and programmable,
lead configurations. Pentium-based, NT technology
platform and hard drive storage delivers local
and Catalyst MUSE database access.
GE Marquette Case Signal Processing:
ECG analysis frequency: 500Hz; ST measurements:
ST amplitudes, slope, integral, index, ST/HR
slope, ST/HR loops, ST/HR index up to 15 leads;
E, I and post-I point: Manual or computer selected;
Signal processing technique: Incremental median
updating; Baseline correction: Cubic Spline
or Finite Residual Filter (FRF) algorithm;
QRS detection and analysis: Based on automatic
or manual lead selection; ECG output: Real-time
ECG/QRS beep/TTL synchronization output; Heart
rate: Automatic arrhythmia detection, documentation
and annotation; Full disclosure ECG: Beat-to-beat
record and event review; Reanalysis: Post-test
medians re-measurement from E, I, post-I point
selections; ECG interpretation: (Optional)
12SL adult and pediatric ECG analysis program;
Additional ECG function: Vectorcardiography.
GE Marquette Case Data Acquisition:
Technology: Active, "Type BF" floating
isolated powered 14 channel acquisition module
with built-in leadfail detection and lead prep
impedance measurement; Sampling rate: Over-sampling
@ 4000 Hz, 12 leads; Dynamic range: 320 mV,
+ 10mV signal superimposed on + 150mV DC offset;
Resolution: 4.88 u V/LSB @ 500 Hz; Noise: <15u
V peak-to-peak noise over 0.01 to 150 Hr (-3
dB) bandwidth; Frequency response: -3 dB, display
and writer High pass filter: 0.01 (or 0.05
Hz, special use) with DC offset control; Low
pass filter: 20, 40, 100, 150 Hz (selectable);
Line filter: 50.0 or 60.0 Hz notch filter (selectable);
Baseline correction: Cubic spline algorithm;
Artifact/baseline correction: FRF algorithm;
Common mode rejection:> 140 dB (123 dB with
AC filter disabled); Input impedance:> 10MW
@ 10 Hz, defibrillator protected; Patient leakage: <10uA
Pace detect: Orthogonal LA, LL and V6; 750
uV@ 50us
GE Marquette Case Communications/Storage:
Catalyst MUSE CV system compatible via diskette,
network (optional) or modem (optional) MUSE
CV Web compatible for retrieval view and printing
of Catalyst MUSE CV system data Remote view
(optional) of real-time stress test and remote
location
Local storage: Minimum 3.2 Gb hard drive storage
of complete ECG record and test results
Local network storage (optional)
GE Marquette Case Display
Display type: High-resolution,
color 0.25 mm dot pitch raster-written CRT;
Display resolution: 1024 x
768 x 70 Hz refresh;
Display size: 15" (standard) or 17" (optional)
diagonal SVGA; Monitored leads: 3, 6 or 12;
Displayed leads: Number or screen 3 or 6 (12 lead check);
Display format: 3 rhythm + medians, 3 rhythm + trends, 6 rhythm,
4 x 2.5 + 1 rhythm, 2 x 6;
Display speeds: 25, 50 mm/s; Display sensitivity/gain: 2.5,
5, 10, 20 mm/mV;
Displayed vital signs data (configurable): Heart rate, target
heart rate, blood pressure, exercise clock, stage clock, phase clock, protocol,
speed, grade, Watts, METS, RPP and Sp02;
Displayed data: ST scan/median complexes, arrhythmias, ventricular
ectopic/min counter, 3 to 15 waveforms, lead check torso and 12 leads, waterfall
display, trends, tabular summary, stored ECG strips, interpretation, time-of-day
clock, patient name, warning messages and prompts
GE Marquette Case Write
Writer Technology: "Instant" load,
thermal dot array;
Writer leads: 3,6,12 or 15 leads (standard, NEHB, Cabrera,
configurable);
Writer speeds: 5, 12.5 25 and 50 mm(sec (+2%);
Writer sensitivity/gain: 2.5, 5, 10 or 20 mm/mV (+5%);
Writer resolution: Horizontal 1000 lines/sec x 200 dpi dedicated
local printing.
200 x 200 dpi generic printing;
Paper type: Thermal, perforated,
fan fold, 300 sheets/pack;
Paper size: A size 214.63 x 280 mm (8.45 x
11 in) A4 size 210 x 297,5 mm (8.27 x 11.7
in)
GE Marquette Case Power Requirements
Power supply: AC operation
only; Operation voltage range: 90-135 VAC,
45-63 Hz, 2.8 A; 180-270 VAC, 45-63 Hz, 1.4
A; Power consumption: 350 W max (1200 BTU/Hr) <250 W normal (850 BTU/Hr) <30
W standby (100 BTU/Hr)
GE Marquette Case Physical Specifications
Height: 160 cm (63 in) with 17" monitor
installed;
Width: 76 cm (30 in) with paper tray removed
104 cm (41 in) with paper tray;
Depth: 86.5 cm (34 in); Weight: 68 kg (150
lb) without monitor 86 kg (190 lb) with 17" monitor
Interfaces included
|